Light RFP
Federal · SAM.gov · Bethesda, MD

Tokai Hit STXG-GSI2X stage top incubator for Leica Stellaris STED microscope

This opportunity closed on Jul 25, 2026. Browse current bids below.
Closed Jul 25, 2026·General ConstructionView official posting ↗
Issuing agency
Department of Health and Human Services
Method
Combined Synopsis/Solicitation
Category
NAICS 334516
PIN
75N98026Q00794
Location
Bethesda, MD
Posted
Jul 15, 2026
Takeaways by LightRFP AI

Generate a quick AI overview of this bid, pulled from its details and the attached documents.

Description

Requirements:

Provides accurate dynamic control of temperature, humidity, CO2 and oxygen to maintain required cell culture conditions for live cell confocal and STED super-resolution imaging. The stage top incubator is required in order to perform live cell imaging with the unique resolution capabilities of our Leica Stellaris 5 STED microscope. It currently does not have a stage top incubator, and therefore, is unable to perform this type of imaging requested by NCI researchers.

Stage top incubator controller must have two feedback modes, chamber temperature or specimen temperature, while it also continuously monitors the room temperature to maintain specimen temperature within set point anywhere between 25 – 50 degrees Celsius. The built-in digital gas mixer controller accepts 100% CO2 from lab supply to regulate CO2 concentration at the sample, with a CO2 range of 5.0 to 20% with real-time sample feedback. In addition, humidity is actively controlled with an internal humidifier to sense sample chamber humidity. The electrically heated stage top incubator chamber must have magnets for easy interchange of specimen holders, including inserts for 35 mm coverglass dish, 50/60mm coverglass dish, chamber slide, coverglass chamber, and open frame insert for imaging multi-well chambers. A flexible heated collar that accommodates a range of objective lens diameters must also be included.

The components of the stage top incubator must include:

  • Tokai Hit STXG-GSI2X-SET Stage Top Incubator for Super Z galvo stage (inverted microscope stands)

  • Pure (100%CO2) gas use (110- 240V) for individual temperature settings to create optimum temperature and humidity conditions without difficulty for long term live cell imaging.

  • Setting temperature range from 25 to 50°C.

  • Dimensions of 112 x 87 mm compatible with Leica Super Z galvo stage on existing Leica Stellaris 5 STED microscope.

  • the lightest chamber (less than 140g) for Super Z Glavo stage

  • temperature controller with built-in digital gas mixer (CO2: 5.0 - 20.0%)

  • Objective heater

  • Magnetic Dish holder and fixing lid for 35mm, 50/60mm dish, chamber slide, slide glass and chamber slide use.

  • External Sensor, enable to autoclave repeatability for Sample Feedback regulation.

  • Tokai Hit original data logging, temp/CO2 program control, software compatible to use with LAS-X software

Documents1
Requirement 33.docxDOCX

Contact

Name
John Burton
Address
Bethesda, MD 20892
Full source recordfrom SAM.gov
Type
Combined Synopsis/Solicitation
Title
Tokai Hit STXG-GSI2X stage top incubator for Leica Stellaris STED microscope
Active
Yes
Base Type
Combined Synopsis/Solicitation
Notice Id
7f0309dced09447697a2aca2279013b4
Naics Code
334516
Department
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Posted Date
Jul 15, 2026
Categorized By
llm-nc
Point Of Contact
Type
primary
Email
burtonjo@mail.nih.gov
Full Name
John Burton
Description Text
Requirements: Provides accurate dynamic control of temperature, humidity, CO2 and oxygen to maintain required cell culture conditions for live cell confocal and STED super-resolution imaging. The stage top incubator is required in order to perform live cell imaging with the unique resolution capabilities of our Leica Stellaris 5 STED microscope. It currently does not have a stage top incubator, and therefore, is unable to perform this type of imaging requested by NCI researchers. Stage top incubator controller must have two feedback modes, chamber temperature or specimen temperature, while it also continuously monitors the room temperature to maintain specimen temperature within set point anywhere between 25 – 50 degrees Celsius. The built-in digital gas mixer controller accepts 100% CO2 from lab supply to regulate CO2 concentration at the sample, with a CO2 range of 5.0 to 20% with real-time sample feedback. In addition, humidity is actively controlled with an internal humidifier to sense sample chamber humidity. The electrically heated stage top incubator chamber must have magnets for easy interchange of specimen holders, including inserts for 35 mm coverglass dish, 50/60mm coverglass dish, chamber slide, coverglass chamber, and open frame insert for imaging multi-well chambers. A flexible heated collar that accommodates a range of objective lens diameters must also be included. The components of the stage top incubator must include: - Tokai Hit STXG-GSI2X-SET Stage Top Incubator for Super Z galvo stage (inverted microscope stands) - Pure (100%CO2) gas use (110- 240V) for individual temperature settings to create optimum temperature and humidity conditions without difficulty for long term live cell imaging. - Setting temperature range from 25 to 50°C. - Dimensions of 112 x 87 mm compatible with Leica Super Z galvo stage on existing Leica Stellaris 5 STED microscope. - the lightest chamber (less than 140g) for Super Z Glavo stage - temperature controller with built-in digital gas mixer (CO2: 5.0 - 20.0%) - Objective heater - Magnetic Dish holder and fixing lid for 35mm, 50/60mm dish, chamber slide, slide glass and chamber slide use. - External Sensor, enable to autoclave repeatability for Sample Feedback regulation. - Tokai Hit original data logging, temp/CO2 program control, software compatible to use with LAS-X software
Response Dead Line
Jul 25, 2026
Full Parent Path Name
DEPARTMENT OF HEALTH AND HUMAN SERVICES.NATIONAL INSTITUTES OF HEALTH.NATIONAL INSTITUTES OF HEALTH OLAO